IC-10, Ion Chamber, Wide-range Detector for Gamma, X-ray and Beta Radiation

Item number: BAK-1230

  • External sensor for ROTEMradiation meter RAM DA-3-2000
  • Rugged ion chamber detector
  • Fixed area monitoring of gamma, X-ray and beta exposition
  • Measuring range: 1Sv/hto 250mSv/h
  • Aluminium housing
  • Built-in calibration factor
  • Optional connection cables to the meter with up to 10m length

Category: Radioactivity


3.671,15 €

including19% VAT.

: 0 workdays

pc.


Description

IC-10, Ion Chamber, Wide-range Detector for Gamma, X-ray and Beta Radiation

The IC-10 is a rugged ion chamber based detector designed for accurate and highly stable field measurements. The detector is used for measuring exposure rates of gamma, X-ray, and beta radiation over a wide range.
The IC-10 Detector, used with the RAM DA-3-2000makes an excellent choice for maintaining the highest safety standards in a wide variety of applications including, radiochemistry, nuclear medicine, transport and storage, radiotherapy, research laboratories and health physics.
Applications:
  • Fixed area monitoring for cyclotrons and accelerators
  • Medium radiation measuring detector
  • Used in high range areas to measure dose rates and accumulated dose
Features:
  • Sturdy Casing (aluminum)
  • Built-in calibration factor
  • Quick Connect to cable
Technical data:
Sensor type: Ion chamber
Measuring range: 1Sv/h to 250mSv/h (0.1mR/h to 25R/h)
Display range: 0.1Sv/h to 250mSv/h (0.1mR/h to 25R/h)
Gamma sensitivity: 0.04pA/mR/h
Accuracy: 15% of reading within measuring range
Gamma energy dependence: Better than 20% at 20keV to 1.2MeV (137Cs)
Angular dependence: Less than 5% (for 120 of front direction (137Cs)
Ion chamber volume: 500cc
Chamber wall and cover thickness: 300mg/cm (tissue equivalent)
End window thickness: 7mg/cm"
Response time (Scale, Time, Range):
0.1 Sv/h to 10Sv/h, 12s, low
10Sv/h to 100Sv/h, 8s, low
100Sv/h to 2mSv/h, 4s, low
2mSv/h to 250mSv/h, 2s, high
Input signals: Offset/operation - control- High/low - range control
Output signals: TTL pulses, detector status: ID, chamber voltage, malfunction
Temperature range:
Operation: -10C to +50C (15F to 122F)
Storage: -20C to +60C (-5F to 140F)
Humidity range: 40% to 95% RH (non-condensing)
Dimensions:
Chamber: 8.9cm x 11.7cm (3.5" x 4.6")
Handle: 5.2cm x 14.3cm (2" x 5.6")
Weight: 470g
Chamber material: Polypropylene
Handle material: Aluminium, splash-proof
Max. cable length: 10m
One of the optional cables is required to connect the probe to the meter:
BAK-0420 1.5m straight connection cable
BAK-0490 1.5m coiled connection cable