Item number: GMH5430

  • Measurement of conductivity, resistance, salinity, TDS
  • Large double display with background illumination
  • Automatic cell correction with reference solutions
  • incl. calibration protocol
  • Category: Salinity Meters, Nutrient Testers & Redox Transducers


    346,29 €

    including 19% VAT.

    Exp. shipping time: ca. 2 weeks

    : 0 workdays

    pc.


    Description

    GMH 5430, Conductivity Meter, Waterprooft, without Probe

    • Measurement of conductivity, resistance, salinity, TDS
    • Large double display with background illumination
    • Automatic cell correction with reference solutions
    • incl. calibration protocol

    Application:
    Mobile use for:
    + industry and craft
    + measurements of waters and aquaristics, fi sh farming
    + drinking water monitoring, process control, soil measurements
    + food production and control
    + quality management
    Additional applications at laboratory:
    + medicine, pharmacy, chemistry


    Further functions:
    Cell correction
    Manually or automatically with reference solution

    Automatic temperature compensation:
    As conductivity depends strongly on temperature, each conductivity value is only valid at the corresponding temperature. Therefore the device supports temperature compensation, i.e. referring the conductivity to a reference temperature (selectable: 20 °C or 25 °C).

    Supported types of compensation:
    - nLF: Non-linear function of natural waters acc. to DIN EN 27888 (ISO 7888) (Reference temperature 25 °C)
    - Lin: adjustable linear compensation
    - off: no compensation

    Salinity measurement:
    Salinity means the sum of the concentrations of all dissolved salts in water.
    The unit is g/kg. (equals PSU = Practical Salinity Unit).

    TDS measurement (total dissolved solids)
    TDS means the mass concentration of dissolved media in a liquid. The unit is mg/l.

    GLP (Good Laboratory Practice)
    adjustable calibration intervals


    Technical data:
    Measuring range:

        Number of meas. ranges: 5
        Smallest range: 0.000 ... 5.000µS/cm (cell constant 0.01/cm) 
                           or 0.0 ... 500.0µS/cm (cell constant 0.1 ... 1.2/cm (standard))
        Biggest range: 0 ... 5000µS/cm (cell constant 0.01/cm) 
                          or 0 ... 1000mS/cm (cell constant 0.1 ... 1.2/cm (standard))
        Resistivity: 0.005 ... 500.0kOhm * cm (depends on cell constant)
        TDS: 0 ... 5000mg/l (depends on cell constant)
        Salinity: 0.0 ... 70.0 (g salt / kg water)
        Temperature: -5.0 ... +100.0°C, Pt1000 or NTC (10k)
    Supported cell constants: 4.000 ... 15.000/cm; 0.4000 ... 1.5000/cm; 0.04000 ... 0.15000/cm;  0.004000 ... 0.015000/cm
    Accuracy (at nominal temperature = 25°C):
        Conductivity: ±0.5% of m.v. ±0.1 % FS (depends on electrode)
        Temperature: ±0.2K
    Connection:
        Conductivity, temperature: 1 x 7-pole bayonet connector for connection of different measuring cells, supported temperature sensors: Pt1000 or NTC (10k)
        Interface / ext. supply: 4-pole bayonet connector for serial interface and supply (with accessory: USB adapter USB 5100)
    Display: 4 ½ digit 7-segment, illuminated (white)
    Operating conditions: Device: -25 ... +50°C, 0 ... 95% RH (non-condensing)
    Storage temperature: -25 ... +70°C
    Background illumination: duration adjustable (off, 5s ... 2min.)
    Power supply: 2 x AAA battery (included), power consumption 6.25mA
        Battery life: approx. 160h (without background illumination)
    Protection class: IP65/IP67
    Housing: Impact-resistant ABS plastic housing, integrated pop-up clip
    Dimensions: 160 x 86 x 37mm³ (H x W x D) incl. silicone protection cover
    Weight: approx. 250g incl. battery and protection cover
    Scope of supply: Device, K 50 BL, battery, calibration protocol, manual
                                The instrument is supplied without a probe. This needs to be ordered separately.


    Optional accessories:
    Conductivity electrodes: LF 200 RW, LF 210, LF 400 or LF 425
    GNG 5/5000  Plug-in power supply 5VDC, suitable for devices of the series GMH 5000
    GKK 3500     Device case (394 x 294 x 106mm³) with eggcrate foam and cut-outs for 1 device
    Conductivity control solutions: GKL 100: 1413µS/cm; GKL 101: 84µS/cm; GKL 102: 50mS/cm
    GEH 1           Electroce holder for measuring electrodes with plastic handle