TA049, 500 MHz Scope Probe, x 10 This is a passive high-impedance oscilloscope probe designed and calibrated for use on instruments having an input impedance of 1 MΩ shunted by 15 pF. However, it may be compensated for use with instruments having an input capacitance of 10 to 35 pF. The probe has a further three trimmers for high-frequency compensation adjustment.
Technical data:
| Bandwidth |
DC to 500 MHz |
| Attenuation |
x10 |
| Rise time |
0.7 ns |
| Input resistance |
10 MΩ when used with oscilloscopes with 1 MΩ input |
| Input capacitance |
approximately 11 pF |
| Compensation range |
10 to 35 pF |
| Working voltage |
Maximum 600 V (DC + Peak AC) |
| Safety |
Conforms to IEC-1010 |
| Cable length |
1.5 M (approx 5 ft) |
| Weight |
146 g (aprrox 5.2 oz) including accessories and wallet |
| Operating temperature |
0 °C to 50 °C, 80% RH |
| Storage temperature |
0 °C to 50 °C, 80% RH |